Diagnostic device > Diagnostic agglutination antisera
ANTISERA FOR SLIDE AGGLUTINATION       Code: AS 415
 
ESCHERICHIA COLI O:157 H7
 
Purpose and usage:
Antiserum ESCHERICHIA COLI O:157 H7 is designed for the rapid and easy determination of  serological group of E. coli O:157 H7.

Technique:
1. Put one drop of  the steril saline on a slide and resuspend in it  one  colony of the tested strain of E. coli (use 24hours culture).
2. Next to this drop put one drop of  the antiserum and mix well  both drops.
3. Rock the slide for 1 minute by circling movements.

 
Reading the results and evaluation of the reaction:
Perform the  reading against dark background:
- forming of agglutinates and clearing the surrounding  liquid: +    positive reaction
- agglutinates do not form and the liquid remains turbid:        -     negative reaction     
 
Evaluation of the results:
1. Positive reaction – tested strain of E. coli  is serologic group O:157 H7
2. Negative reaction – tested strain of E. coli is from other serological group
 
NOTE: If the agglutination of tested strain in saline occurs, the tested strain is spontaneously agglutinating and it is unsuitable for slide agglutination testing.
 
Warning:
1. Antiserum is protected with sodium azide. Nevertheless it is necessary to avoid its contamination during its use (you must not touch the slide with the dropper during dropping the serum!).
2. Antiserum is delivered in liquid form 1 ml or 2 ml and  is designed for laboratory use only.
3. Antiserum should be kept in a dark place at +4°C and should not be used beyond the expiration date shown on the packing.

 